New TopShim run:

version = topshim 2.4.0
date = 2026-11-02
time = 17:27:57.094
command = topshim  convcomp ordmax=4

FINE SHIMMING

Parameters:

probe = Z104275_0310
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Low S/N encountered. The acquisition parameters will be changed to improve this issue.
Iteration  1: FWHM Before= 0.7 Hz, Target= 0.3 Hz, After= 0.2 Hz
Iteration  2: FWHM Before= 0.2 Hz, Target= 5.4 Hz, After= 3.2 Hz
Iteration  3: FWHM Before= 3.2 Hz, Target= 0.3 Hz, After= 2.6 Hz
Iteration  4: FWHM Before= 2.6 Hz, Target= 0.6 Hz, After= 1.6 Hz
Iteration  5: FWHM Before= 1.6 Hz, Target= 0.4 Hz, After= 0.4 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.741

shim changes: 
Z	+3929
Z2	+7935
Z3	+17698
Z4	+23815

duration = 3 min 34 sec
completed successfully
finished Mon Nov 02 17:31:29 2026


========================================================================