New TopShim run:

version = topshim 2.4.0
date = 2026-11-06
time = 18:11:50.614
command = topshim  convcomp ordmax=4

FINE SHIMMING

Parameters:

probe = Z104275_0310
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Low S/N encountered. The acquisition parameters will be changed to improve this issue.
Iteration  1: FWHM Before= 1.1 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  2: FWHM Before= 0.4 Hz, Target= 0.4 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.731

shim changes: 
Z	+29
Z2	+13
Z3	-118
Z4	+171

duration = 1 min 47 sec
completed successfully
finished Fri Nov 06 18:13:36 2026


========================================================================