New TopShim run:

version = topshim 2.4.0
date = 2026-09-28
time = 16:51:16.393
command = topshim  convcomp ordmax=4

FINE SHIMMING

Parameters:

probe = Z104275_0310
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Low S/N encountered. The acquisition parameters will be changed to improve this issue.
echo time must be reduced
echo time must be reduced

NOTE: very poor map quality encountered despite the CONVCOMP option being used.

Required SINO has not been achieved!
Please increase DURMAX if a better map quality is needed.
Required SINO has not been achieved!
Please increase DURMAX if a better map quality is needed.
Iteration  1: FWHM Before=98.7 Hz, Target= 0.2 Hz, After= 0.5 Hz
Iteration  2: FWHM Before= 0.5 Hz, Target= 1.5 Hz, After= 1.3 Hz
Iteration  3: FWHM Before= 1.3 Hz, Target= 2.2 Hz, After= 2.2 Hz
Iteration  4: FWHM Before= 2.2 Hz, Target= 1.1 Hz, After= 2.7 Hz
Iteration  5: FWHM Before= 2.7 Hz, Target= 0.9 Hz, After= 3.4 Hz
Iteration  6: FWHM Before= 3.4 Hz, Target= 0.3 Hz, After= 1.9 Hz
Iteration  7: FWHM Before= 1.9 Hz, Target= 0.3 Hz, After= 0.2 Hz
Iteration  8: FWHM Before= 0.2 Hz, Target= 0.7 Hz, After= 0.5 Hz
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 1.01

shim changes: 
Z	-13094
Z2	-19357
Z3	-24784
Z4	-18330

duration = 12 min 53 sec
completed successfully